SPECIFICATIONS
Performance
Electrical
Environmental Conditions
Environmental Protection
Physical
PERFORMANCE
Measurement Range
SpO2
0–100%
Pulse Rate
20–250 beats per minute (bpm)
Accuracy
SpO2
Adults 70–100% ±2 digits
0–69% unspecified
Neonates 70–100% ±3 digits
0–69% unspecified
Note: Accuracies are expressed as plus or minus “X” digits
(oxygen saturation percentage points) between
saturations of 70–100%. This variation equals plus or
minus one standard deviation (1SD), which encompasses
68% of the population. All accuracy specifications are
based on testing the subject monitor on healthy adult
volunteers in induced hypoxia studies across the specified
range. Adult accuracy is determined with Oxisensor II
D-25 sensors. Neonatal accuracy is determined with
Oxisensor II N-25 sensors. In addition, the neonatal
accuracy specification is adjusted to take into account the
theoretical effect of fetal hemoglobin in neonatal blood
on oximetry measurements.
53