Setup
MONITOR SETUP
General
WARNING: To ensure patient safety, do not place the
monitor in any position that might cause it to fall on the
patient.
WARNING: As with all medical equipment, carefully route
patient cabling to reduce the possibility of patient
entanglement or strangulation.
WARNING: Disconnect the NPB-290 and Nellcor sensor
from the patient during magnetic resonance imaging (MRI)
scanning. Induced current could potentially cause burns.
The NPB-290 may affect the MRI image; the MRI unit may
affect the accuracy of oximetry measurements.
WARNING: To ensure accurate performance and prevent
device failure, do not subject the NPB-290 to extreme
moisture, such as direct exposure to rain. Such exposure may
cause inaccurate performance or device failure.
WARNING: Do not use an NPB-290 monitor, sensor, cables,
or connectors that appear to be damaged.
WARNING: The NPB-290 is not defibrillator-proof.
However, it may remain attached to the patient during
defibrillation or while an electrosurgical unit is in use, but the
readings may be inaccurate during use and shortly thereafter.
Setting Up The NPB-290
The NPB-290 operates on AC power when the hospital-grade
power cord is connected to both the monitor and an AC power
source (wall outlet).
16