Start-Up and Use
Viewing Current Alarm Silence Duration
To view the current audible alarm silence duration, press and hold
the Alarm Silence button for less than 3 seconds. The display
will show the current duration in seconds, as shown:
To adjust the audible alarm silence duration, while holding down
the Alarm Silence button, use the Adjust Up or Adjust Down
button to set the duration to 30, 60, 90, or 120 seconds, or OFF
(Alarm Silence mode). While the alarm is silenced temporarily,
the Alarm Silence indicator lights during the silence duration
period.
Visual indications of an alarm condition are not affected by
silencing the audible alarms.
Disabling Audible Alarms
Setting the alarm silence duration to OFF means that no audible
alarms will be produced by the monitor.
To set the alarm silence duration to OFF, press and hold the
Alarm Silence button for less than 3 seconds and use the Adjust
Up button to increase the current setting until “OFF” is displayed.
The next time the Alarm Silence button is pressed, the Alarm
Silence indicator flashes, indicating that audible alarms have been
disabled.
Visual indications of an alarm condition are not affected by
disabling the audible alarms.
The ability to set the alarm silence duration to OFF can be
enabled or disabled by qualified service personnel as described in
the service manual.
The factory default is that the capability of setting the alarm
silence duration to OFF is enabled.