Sensors and Accessories
Inaccurate measurements can be caused by:
incorrect application of the sensor
placement of the sensor on an extremity with a blood
pressure cuff, arterial catheter, or intravascular line
ambient light
patient movement
Loss of pulse signal can occur for the following reasons:
the sensor is too tight
a blood pressure cuff is inflated on the same extremity as
the one with the sensor attached
there is arterial occlusion proximal to the sensor
Use only Nellcor sensors and sensor cables. Select an appropriate
sensor, apply it as directed, and observe all warnings and cautions
presented in the directions for use accompanying the sensor.
Clean and remove any substances such as nail polish from the
application site. Periodically check to see that the sensor remains
properly positioned on the patient.
WARNING: Tissue damage can be caused by incorrect
application or duration of use of an SpO2 sensor. Inspect the
sensor site as directed in the sensor directions for use.
High ambient light sources such as surgical lights (especially
those with a xenon light source), bilirubin lamps, fluorescent
lights, infrared heating lamps, and direct sunlight can interfere
with the performance of an SpO
2 sensor. To prevent interference
from ambient light, ensure that the sensor is properly applied, and
cover the sensor site with opaque material.
Note: Failure to take this precaution in high ambient light
conditions may result in inaccurate measurements.
If patient movement presents a problem, try one or more of the
following remedies to correct the problem.
Verify that the sensor is properly and securely applied.
Move the sensor to a less active site.
Use an adhesive sensor that tolerates some patient motion.
23