Mallinckrodt NPB-190 Manual del operador Pagina 39

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 79
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 38
Start-Up and Use
At Initial Power-Up
Immediately after the NPB-290 displays its software version
numbers and POST is completed, the monitor enters Pulse Search
(the Pulse Search indicator lights), if a sensor is properly
connected to the monitor. During this Pulse Search, the display
reads zeroes while the monitor searches for a valid pulse. When a
valid pulse is detected, the NPB-290 begins monitoring.
After Taking Measurements
If a pulse was previously acquired and then lost, the NPB-290
enters Pulse Search, and the Pulse Search indicator lights. The
display flashes the last detected readings while the monitor
searches for a valid pulse. When the monitor considers the pulse
“lost,” it displays flashing zeroes and a high priority alarm
sounds.
When a valid pulse is again detected, the NPB-290 exits pulse
search and displays the current readings. The Pulse Search
indicator goes out.
Refer to the “Management of the Loss-of-Pulse Alarm” paragraph
later in this section for a detailed description of Pulse Search,
motion, and the loss-of-pulse alarm.
Automatic Shutdown
The monitor automatically shuts off after 15 minutes, provided all
of the following conditions are in effect:
the monitor is running on its internal battery
no alarms are present (other than low battery or a
non-correctable error)
no buttons have been pressed, and
no pulse has been detected
Once a valid pulse is again acquired (within 15 minutes), the
instrument begins monitoring again.
31
Vista de pagina 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 78 79

Comentarios a estos manuales

Sin comentarios