Controls, Indicators, and Symbols
The Pulse Search indicator. Lights continuously prior to
initial acquisition of a pulse signal, and during Pulse
Search mode. It flashes during loss of pulse.
DESCRIPTION OF VISUAL INDICATORS AND DISPLAYS
The %SpO2 Display. Shows the hemoglobin oxygen
saturation level.
The Pulse Amplitude indicator. A 10-segment display
that shows the relative pulse amplitude. As the detected
pulse becomes stronger, more contiguous segments light
with each pulse.
The digital Pulse Rate Display. Shows the pulse rate in
beats per minute.
DESCRIPTION OF AUDIBLE INDICATORS
Following are descriptions of NPB-290 audible indicators.
Power-On Self-Test (POST)
pass
One 1-second tone indicating that
the NPB-290 has been turned on
and successfully completed the
power-on self-test.
Valid Button Press A single tone indicating that an
appropriate button has been
pressed.
Invalid Button Press A single low-pitched tone
indicating that a button has been
pressed that is not appropriate for
the current state of the monitor.
High Priority Alarm A high-pitched, fast, pulsing tone
indicating loss-of pulse.
Medium Priority Alarm A medium-pitched, pulsing tone
indicating an SpO2 or pulse rate
limit violation.
Low Priority Alarm A low-pitched, slow, pulsing tone
indicating a sensor disconnect,
low battery, or monitor failure.
8