Start-Up and Use
the numbers and have them available prior to
requesting technical assistance.
5. If the NPB-290 detects an internal problem during the self-
test, an error tone sounds and the monitor displays an Error
Code and corresponding number (see the Troubleshooting
section).
6 Upon successful completion of the self-test, the NPB-290
sounds a 1-second tone indicating that the monitor has passed
the test.
WARNING: If you do not hear the POST pass tone, do not
use the monitor.
In addition to serving as the POST pass verification, the
POST pass tone also functions as audible confirmation that
the speaker is performing properly. If the speaker does not
function, the alarms cannot sound.
If a sensor is connected to the monitor and the patient, the
Pulse Search indicator lights and the NPB-290 displays
zeroes in its SpO
2 and pulse rate fields while it searches for a
valid pulse from a patient. If no sensor is attached, dashes are
displayed and the Pulse Search indicator is not lit.
Silencing or Adjusting Pulse Beep Volume
The pulse beep volume can be adjusted within a range of 0-10.
At Level 0, the pulse beep volume is silenced; at Level 10, the
volume is at its highest setting.
This volume can be adjusted by pressing either the Adjust Up or
Adjust Down button located on the right side of the front panel.
Note: The default volume setting can be changed by qualified
service personnel by following the instructions described
in the NPB-290 service manual.