Start-Up and Use
Ensure that the voltage selector switch on the rear panel is set to
the proper voltage.
Power-On Self-Test (POST)
WARNING: Ensure that the speaker is clear of any
obstruction. Failure to do so could result in an inaudible
alarm tone.
1. Plug an appropriate Nellcor sensor firmly into the MC-10 or
SCP-10 and lower the MC-10 or SCP-10 sensor lock over the
sensor connector until it clicks into place. Insert the MC-10
or SCP-10 into the NPB-290 sensor port. Apply the sensor to
the patient as described in the sensor directions for use.
2. Turn on the NPB-290 by pressing the Power On/Off button.
After a few seconds, the monitor automatically starts a
power-on self-test (POST), which tests its circuitry.
Caution: During POST (immediately after power-up),
confirm that all display segments and indicators light.
3. While performing POST, the self-test display appears for
approximately 2 seconds. During this time:
• All indicator lights illuminate
• All segments of all numeric digits light and change
from red to green
• All segments of the Pulse Amplitude Display light
Caution: If any indicator or display element does not light,
do not use the monitor. Instead, contact qualified service
personnel or your local Mallinckrodt representative.
4. Once the display test portion of POST is complete, the
NPB-290 software version is displayed for approximately 2
seconds.
Note: Software version numbers are often needed when
calling Mallinckrodt’s Technical Services
Department or your local Mallinckrodt
representative for technical assistance. Write down