Setup
The supply voltage selector switch allows connection of the
monitor to AC power ranging from 100 VAC to 240 VAC. The
switch has two positions: one for 100-120 VAC (“115”), and one
for 200-240 VAC (“230”). Ensure that the supply voltage
selector switch on the rear panel is set to the proper voltage.
Operating on a discharged battery
The NPB-290 will not operate when its internal battery is
discharged, even when the monitor is connected to AC power.
Instead, the error code “EEE 04” is displayed. This feature
prevents the accidental use of the monitor with a dead battery.
The monitor is only capable of indicating a loss of AC power if
its internal battery is functional.
The battery may discharge during prolonged storage or shipment.
If the monitor has been in storage for more than 2 months, it is
important to plug the monitor into an AC outlet and allow the
battery to charge for approximately 30 minutes before attempting
to operate the instrument on AC power.
To fully charge a dead battery, plug the monitor into an AC wall
outlet and allow it to charge for 14 hours.
If transporting the monitor, or if AC power is not available, you
may operate the NPB-290 on battery power for a limited amount
of time. In that case, skip Steps 2 and 3 below.
1. Place the NPB-290 on a flat surface near the patient. With an
optional mounting plate available from Mallinckrodt, the
monitor may be mounted to a wall.
2. Plug the female connector end of the power cord into the rear
of the monitor as shown in Figure 3. Use only the hospital-
grade power cord provided by Mallinckrodt.
17