1
SAFETY INFORMATION
General Safety Information
GENERAL SAFETY INFORMATION
This section contains important safety information related to
general use of the N-395 pulse oximeter. Other important safety
information appears throughout the manual in sections that relate
specifically to the precautionary information. Read all text
surrounding all precautionary information.
Important! Before use, carefully read this manual, accessory
directions for use, all precautionary information in boldface type,
and specifications.
WARNING: Explosion hazard. Do not use the N-395 pulse
oximeter in the presence of flammable anesthetics or gases.
WARNING: The N-395 is a prescription device and is to be
operated by qualified personnel only.
WARNING: Pulse oximetry readings and pulse signal can be
affected by certain ambient environmental conditions, sensor
application errors, and certain patient conditions. See the
appropriate sections of the manual for specific safety
information.
WARNING: Chemicals from a broken LCD display panel
are toxic when ingested. Use caution when handling a
monitor with a broken display panel.
Caution: When connecting the N-395 to any instrument,
verify proper operation before clinical use. Both the N-395
and the instrument connected to it must be connected to a
grounded outlet. Accessory equipment connected to the
monitor’s data interface must be certified according to IEC
Standard 950 for data-processing equipment or IEC
Standard 601-1 for electromedical equipment. All
Comentarios a estos manuales