Data Port Protocol
95
N-395 VERSION 2.0.0.0 TREND SpO2 Limit: 70-100% PR Limit: 60-160BPM
TIME %SpO2 BPM PA Status
12-NOV-99 14:00:05 100 120 150
12-NOV-99 14:00:07 100 121 154
12-NOV-99 14:00:09 100 120 150
Output Complete
Figure 24: Trend Data Printout (ASCII MODE)
Once a trend printout has begun, it cannot be aborted without
turning off the N-395 or the printer.
TREND DATA PRINTOUT (GRAPH MODE)
The graph mode (Figure 25) disables all printout functions
except trend data. Trend printouts will be graphical if connected
to an approved serial printer. Contact Mallinckrodt’s Technical
Services Department for a list of approved serial printers.
Figure 25: Trend Data Printout (GRAPH MODE)
NURSE CALL
WARNING: The nurse call feature should not be used as the
primary source of alarm notification. The audible and visual
alarms of the monitor, used in conjunction with clinical signs
and symptoms, are the primary source for notifying medical
personnel that an alarm condition exists.
The N-395 provides two different types of nurse call interfaces:
an RS-232 level and solid state relay closure. The solid state
relay-based nurse call function is available when the monitor is
operating either on AC power or when powered by battery.
The remote location will be signaled anytime there is an audible
alarm.
Pin 11 on the data port is the RS-232 level nurse call signal and
pin 10 is ground (see Table 10). The nurse call polarity
(normally high or normally low) and whether the monitor is in