Start-Up and Use
59
Table 6: Factory Default Settings (Adult)
Monitoring Mode Adult
%SpO2 Lower Alarm Limit: 85%
%SpO2 Upper Alarm Limit: 100%
Alarm Silence Duration OFF
Setting:
Enabled
Alarm Silence Duration: 60 seconds
Alarm Silence Reminder: Enabled
Alarm Volume: 75 dB(A) peak at 1 meter
(volume setting of 5)
Data Port Baud Rate: 9600
Data Port Protocol: ASCII
Display Contrast: Midrange
Display Format: Pleth
Language: English
Nurse Call Polarity: Normally Low
Pulse Beep Volume: 72 dB(A) at 1 meter
(volume setting of 4)
Pulse Rate Lower Alarm Limit: 40 beats per minute
Pulse Rate Upper Alarm Limit: 170 beats per minute
SatSeconds Off
Trend Display: %SpO2