Start-Up and Use
57
Datex-Ohmeda Communications
The Datex-Ohmeda monitor AS/3 must be configured for
communications with the Nellcor N-200 monitor in order to
communicate with the N-395 monitor. Refer to the AS/3
operator’s manual for instructions on configuring the AS/3
monitor.
The N-395 sends SpO
2, pulse rate, and alarm status data to the
Datex AS/3 monitor.
The RS-232 hardwire interface cable has a DB-15 connector for
the N-395 and the applicable connector for the Datex monitor.
Nellcor cable part number 902255 is recommended for this
interface.
Note: Spare parts for the N-395 are listed on the Internet,
confirm the cable part number before ordering a cable.
The Internet address is:
http://mallinckrodt.com/respiratory/resp/Serv_Supp/Apa
rtweb/main/PartAcceMenu.html.
Corrupt data will be indicated by a Communications Error
displayed on the Datex monitor. The Datex monitor will detect
corrupt data in less than 11 seconds.
When the N-395 is in the Datex mode of operation the interface
baud rate must be set to 2,400 bits per second. Pressing the
SETUP softkey, then the NEXT softkey, and then the COMM
softkey to select BAUD. Use the ADJUST UP or ADJUST
DOWN buttons to select the correct baud rate.
WARNING: Do not silence the N-395 audible alarm or
decrease its volume if patient safety could be compromised.
The Datex-Ohmeda monitor does not indicate audible or visual
alarms. The N-395 monitor must be able to sound an audible
alarm in order to maintain patient safety.
LANG
The LANG softkey provides selection of the language displayed
on the screen.