
Start-Up and Use
60
Table 7: Factory Default Settings (Neonate)
Monitoring Mode Neonate
Note: There are different default settings for the neonate mode.
%SpO2 Lower Alarm Limit: 80%
%SpO2 Upper Alarm Limit: 95%
Pulse Rate Lower Alarm Limit: 90 beats per minute
Pulse Rate Upper Alarm Limit: 190 beats per minute
SatSeconds:Off
Nurse Call Feature
WARNING: The nurse call feature should not be used as
the primary source of alarm notification. The audible and
visual alarms of the monitor, used in conjunction with
clinical signs and symptoms, are the primary sources for
notifying medical personnel that an alarm condition exists.
The nurse call feature of the N-395 works in conjunction with
the nurse call system of the institution when the monitor sounds
an audible alarm. It is accessed through the data port (pins 7, 8,
10, 11, or 15, as indicated in Table 10).
WARNING: The nurse call feature is not functional
whenever the monitor alarms are silenced.
The nurse call feature is available when the N-395 is operated
on AC power or its internal battery, and the monitor has been
electronically connected to the hospital’s nurse call system.
Qualified service personnel may refer to the N-395 service
manual for complete connection instructions.
Prior to using the monitor in a clinical setting, test the nurse call
feature by creating an alarm condition, then verifying that the
hospital’s nurse call system is activated.
Comentarios a estos manuales