Start-Up and Use
46
Alarm Limits Changed Indicator
If alarm limits are changed from the N-395’s power-on defaults,
a decimal point appears after the displayed value and in the
%SpO
2 and Pulse Rate Display as illustrated in Figure 12. The
decimal point remains on the display until the N-395 is turned
off or the limit is returned to its default value.
Figure 12: Alarm Limits Selection
TREND
The N-395 can graphically display trends for SpO
2, pulse rate,
or both. Trend data is stored at 4-second intervals. When the
TREND softkey is pressed, “READING TRENDS . . ..” is
displayed at the bottom of the screen, indicating that the monitor
is collecting the trend data.
The monitor stores up to 48 hours of trend data. The amount of
trend data displayed on the screen is determined using the
ZOOM softkey. Settings available are 40 seconds, 15 or 30
minutes, and 1, 2, 4, 8, 12, 24, 36, or 48 hours. All data are
displayed in a graph format except the 40-second setting, which
is shown in tabular format.
When the trends are displayed, the most recent readings are on
the right side of the graph. The graph indicates the highest and
lowest parameter values during the period of time represented
by the width of the cursor (vertical dotted line).
The highest and lowest values of the parameter at the cursor are
indicated on the left side of the screen (“95” and “98” in Figure
13). These values are not the current patient readings but
represent the highest/lowest trend values at the cursor.