Data Port Protocol
97
ANALOG OUTPUTS
The N-395 data port also provides analog voltage outputs
between pins 6, 13, 14, and ground (pins 5 or 10), which can be
used to calibrate instruments such as a chart recorder. The
voltage represents a specific measured parameter’s current value.
The voltage differential varies proportionally from 0 to 1 volt as
the pin’s parameter varies over its full range of values, as
indicated in Table 13.
Table 13: Analog Pinouts
Pin Parameter Parameter Range
6%SpO2 0 - 100%
13 Pulse rate 0 - 250 bpm
14 Pleth wave 0 - 255
For example, as the current value of %SpO2 varies from 0 to
100%, the voltage from pin 6 to ground (pin 10) would vary from
0 to 1 volt. A voltage of .94 volts indicates a current %SpO
2
value of 94.
The analog output calibration function can be accessed from the
main menu by pressing SETUP, NEXT, NEXT, ANALOG.
Selecting “0 VOLT” or “1 VOLT” causes that voltage to appear
at pins 6, 3, or 10. Selecting “STEP” causes the voltage to
increase from 0 to 1 volt at 1/10
th
-volt increments, with each step
lasting at least 1 second.
Qualified service personnel using the procedure described in the
N-395 service manual can perform calibration of the analog
output and the attached device.