Troubleshooting and Maintenance
The software version appears in the monitor display each time the
monitor successfully completes the power-on self-test. Write the
number down and have it available whenever requesting technical
assistance.
RETURNING THE NPB-295
Contact Mallinckrodt’s Technical Services Department or your
local Mallinckrodt representative for shipping instructions
including a Returned Goods Authorization number. It is not
necessary to return the sensor. Pack the NPB-295 in its original
shipping carton. If the original carton is not available, use a
suitable carton with appropriate packing material to protect it
during shipping.
Return the NPB-295 by any shipping method that provides proof
of delivery.
MAINTENANCE
Service
The NPB-295 requires no calibration.
The battery should be replaced at least every 24 months. Refer to
the NPB-295 service manual for the battery changing procedure.
If service is necessary, contact qualified service personnel or your
local Mallinckrodt representative.
Periodic Safety Checks
It is recommended that the following checks be performed every
24 months.
Inspect the equipment for mechanical and functional damage.
Inspect the safety relevant labels for legibility.
Performance Verification
If the monitor has been visibly damaged or subjected to
mechanical shock (for example, if dropped), qualified service
personnel should perform the procedure in the Performance
Verification section of the service manual.
55