Start-Up and Use
LIGHT
The LIGHT softkey allows you to turn the backlight on or off.
Default Settings
The NPB-295 is shipped with factory default settings. Default
settings can be changed by authorized technical personnel using
the procedures described in the NPB-295 service manual.
The factory default settings are as follows:
%SpO2 Upper Alarm Limit: 100%
%SpO2 Lower Alarm Limit: 85%
Pulse Rate Upper Alarm Limit: 170 beats per minute
Pulse Rate Lower Alarm Limit: 40 beats per minute
Alarm Volume: 75 dB(A) peak at 1 meter
(volume setting of 5)
Alarm Silence Duration: 60 seconds
Alarm Silence Duration OFF
Setting:
Enabled
Alarm Silence Reminder: Enabled
Pulse Beep Volume: 72 dB(A) at 1 meter
(volume setting of 4)
Data Port Baud Rate: 9600
Data Port Protocol: ASCII
Display Format: Pleth
Trend Display: %SpO2
Display Contrast: Midrange
Language: English
Nurse Call Polarity: Normally Low
44