Setup
Connecting The NPB-295
The NPB-295 operates on AC power when the hospital-grade
power cord is connected to both the monitor and an AC power
source (wall outlet).
The supply voltage selector switch allows connection of the
monitor to AC power ranging from 100 VAC to 240 VAC. The
switch has two positions: one for 100-120 VAC (“115”), and one
for 200-240 VAC (“230”). Ensure that the supply voltage
selector switch on the rear panel is set to the proper voltage.
Operating on a discharged battery
The NPB-295 will not operate when its internal battery is
discharged, even when the monitor is connected to AC power.
Instead, the error code “EEE 04” will be displayed. This feature
prevents the accidental use of the monitor with a dead battery.
The monitor is only capable of indicating a loss of AC power if
its internal battery is functional.
The battery may discharge during prolonged storage or shipment.
If the monitor has been in storage for more than 2 months it is
important to plug the monitor into an AC outlet and allow the
battery to charge for approximately 30 minutes before attempting
to operate the instrument on AC power.
To charge a low battery, connect the monitor to AC power. A
full charge of a completely discharged battery takes 14 hours
while turned off, or 18 hours during regular use.
If AC power is not available, you may operate the NPB-295 on
battery power for a limited amount of time. In that case, skip
Steps 2, 3, and 4 below.
1. Place the NPB-295 on a flat surface near the patient. With
the optional wall mount plate available from Mallinckrodt,
the monitor may be attached to a GCX Poly-mount bracket.
2. Plug the female connector end of the power cord into the rear
of the monitor. Use only the hospital-grade power cord
provided by Mallinckrodt.
13