Start-Up and Use
data port analog voltage.
Press the SETUP softkey once to display VIEW and CLOCK.
Then press NEXT to display COMM and LANG. Press NEXT
again to display NCALL and ANALOG.
VIEW
The VIEW softkey allows you to select the view displayed on the
screen, PLETH or BLIP (magnified). The pleth view displays the
pleth waveform. The BLIP view displays the Pulse Amplitude
Indicator and larger numerical values for easier viewing.
CLOCK
The CLOCK softkey allows you to set the time and date.
Press the SET softkey to access the SELECT softkey. Use the
SELECT softkey to select the item you wish to change. Use the
Adjust Up/Down Buttons to adjust the setting. The date is
expressed as DD-MMM-YY. For example, November 29, 1998,
would be expressed as 29 - NOV - 98.
Note: The NPB-295 will timeout in 5 minutes when the SET
and EXIT softkeys are displayed. However, it will
timeout in 10 seconds after the SET button is pressed if
there are no further button presses.
Press the EXIT button to accept the new settings. Press EXIT
again to return to the previous menu.
COMM
Press the NEXT softkey to access the COMM softkey. The
COMM softkey allows you to select the baud rate and the
protocol of the data port.
After pressing the COMM softkey, use the Adjust Up/Down
Buttons to select a baud rate of 2400, 9600, or 19200. Press Exit
to return to the SETUP submenu.
Press the SELECT softkey to highlight the protocol setting. Use
the Adjust Up/Down Buttons to select the desired setting. ASCII
is used during normal operation and for serial printers. GRAPH
is used to print copies of the NPB-295 trend displays when you
42