Data Port Protocol
89
The pin layouts (as viewed from the rear panel of the N-395) are
illustrated in Figure 22. The conductive shell is connected to
earth ground when connected to a PC or printer.
Figure 22: Data Port Pin Layout
Pins 2, 3, and 5 provide data in RS-232 format.
Pins 1, 4, 9, and 12 provide data in RS-422 format. TxD+ and
TxD- are the differential transmit data pair. RxD+ and RxD- are
the differential receive pair.
No hardware flow control is used. However, in the ASCII mode
XON/XOFF flow control is supported.
BAUD RATE
Pressing the SETUP softkey and then the COMM softkey can
change the baud rate. Use the ADJUST UP/DOWN Buttons to
select a baud rate of 2400, 9600, or 19200, depending on the
capabilities of the attached equipment.
SETTING DATA PORT PROTOCOL
The available data port protocols are:
ASCII
OXINET
CLINICAL
GRAPH
AGILENT – Agilent (HP) communications
SPACELBS – SpaceLabs communications
MARQ – GE Marquette communications
DATEX – Datex-Ohmeda AS/3 communications
Vista de pagina 96